Overview
- Persons in charge: Ran and Ishay
- Past experts: Bo, Ramesh
- Detector location
- wire map:
- HV
- scalers
- ADC
- TDC
- Readme
- Reported problems
Amplitudes alignment
The idea is to align the ADC signals by calibrating the gain & peds DB coefficients. This is done using a script by Yi which fits the signal and peds peaks for each ADC, and then calculate the coefficients for conversion of the amplitude to energy by minimization of chi^2 of the width of (E-p)^2
To run it:
- Step 1 - Produce a pedestal run and a signal run for each spectrometer.
- Step 2 - Run the calibration tool. The script will ask for a pedestal run and a signal run.
It will then fit the pedestal and signal peaks for each ADC and ask for your confirmation
for example:
- Step 3 - Calibration check:
plot E = PR1+PR2 vs p (from vdc):
.

.
Identify the Pion and Electron blobs.
- Step 4 - Install the new DB and estimate resolution by plotting E/p.
Coordinates calibration
HRS-L
HRS-R
Files
- Coda data base
- Data base for pedestals
References
- Tool by Ran (under development)
- Tools by Yi of E04-012
- Bo's calibration tools directory: /w/work5602/transversity/zhaobo/aytest/onlana/test/
- description of RHRS calibration of E97-110